Factory Calibration

The factory calibration laboratory of CETA Testsysteme GmbH is able to perform factory calibrations for the measurand pressure and for the measurand flow.

 

Measurand Pressure

Factory calibrations are carried out using calibrated reference devices, that are calibrated by an accredited laboratory. The traceability of the used measurement devices is documented. The pressure range of -1 bar to +60 bar as gauge pressure and 0.05 bar to 40 bar as absolute pressure is also covered by factory calibrations.  

 

Measurand Flow

The flow measuring range from 0,1 ml/min to 420 l/min with gases can be covered by factory calibrations. Helium leaks can be covered in the range from 1*10-9 to 1*10-6 mbar*l/s. Reference leaks can be calibrated in the range from 5*10-3 to 1*10-6 mbar*l/s. The used calibration standards are recalibrated with an accredited calibration certificate by accredited institutes.

Remark: DAkkS is the abbreviation of “Deutsche Akkreditierungsstelle” (German Accrediting Agency).

 

Data Sheet Download